ISO 80601-2-13:2022 This document is applicable to the basic safety and essential performance of an anaesthetic workstation for administering inhalational anaesthesia whilst continuously attended by a professional operator.
This document specifies particular requirements for a complete anaesthetic workstation and the following anaesthetic workstation components which, although considered as individual devices in their own right, may be utilized, in conjunction with other relevant anaesthetic workstation components, to form an anaesthetic workstation to a given specification:
- anaesthetic gas delivery system;
- anaesthetic breathing system;
- anaesthetic gas scavenging system (AGSS);
- anaesthetic vapour delivery system;
- anaesthetic ventilator;
- monitoring equipment;
- alarm system;
- protection device.
NOTE 1 Monitoring equipment, alarm systems and protection devices are summarized in Table AA.1.
An anaesthetic workstation supplied complete and its individual components are considered as ME equipment or ME systems with regard to the general standard.
NOTE 2 The applicability of this document is indicated in Table AA.2.
This document is also applicable to those accessories intended by their manufacturer to be connected to an anaesthetic workstation where the characteristics of those accessories can affect the basic safety and essential performance of the anaesthetic workstation.
If a clause or subclause is specifically intended to be applicable to anaesthetic workstation components or its accessories only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to an anaesthetic workstation and its individual components including accessories, as relevant.
Hazards inherent in the intended physiological function of an anaesthetic workstation and its individual components including accessories within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 3 See also IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document is not applicable to any anaesthetic workstation intended for use with flammable anaesthetic agents, as determined by Annex BB.

Issue Date

2022-04-29

Category

MED

Included in IECEE System

2023-08-09

Purchase webstore
Test Report Form
Testing Equipment List

No information declared

DISCLAIMER: National differences

The National differences and Group differences, National Deviations and Special National conditions (SNC), are based solely on information provided to the Secretariat by the IECEE Member Bodies and/or NCBs and other sources. The IEC/IECEE is not responsible for, nor will it take any position related to, the accuracy or validity of the information provided. To verify the current status of this type of information, we recommend contacting the Member Body (MB) or National Certification Body (NCB) of the relevant country.

Group Differences are applicable for CENELEC member countries: Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.

Group/Country

Reference

Last Modified

Differences

TRF Addendum

CanadaCanada

2024-08-16

Download55 kb

Id

Name

City

Country

13360

Arnhem

Netherlands

13371

Kista

Sweden

13365

Singapore

Singapore

13342

Helsinki

Finland

13345

Nürnberg

Germany

13427

Singapore 609937

Singapore

13346

München

Germany

13344

Offenbach / Main

Germany

Id

Name

City

Country

13336

Toronto, Ontario

Canada